Aspect Medical Systems
Healthcare ProfessionalsCompany InformationPatientsInvestorsMediaProducts
More Info... Status Messages

Status Messages and Operator Actions

Status Messages:

Causes:

Corrective Actions:

E01
DSC Not Connected

  1. DSC disconnected.
  2. Defective DSC cable.
  3. Defective DSC.
  4. Defective monitor.
  1. Connect DSC. Verify all cable connections.
  2. Inspect / repair cable at connector end.
  3. Replace DSC
  4. Replace monitor

E02
DSC Overcurrent

1.  Defective DSC cable.2.  Defective DSC.3.  Defective monitor.
  1. Replace DSC
  2. Replace DSC.
  3. Replace monitor

E03
DSC Power Regulation

  1. Defective DSC.
  2. Defective DSC cable.
Replace DSC.

E04, E05, E06, E07, E08
DSC Error

  1. Defective DSC
  2. Defective monitor.
  1. Replace DSC.
  2. Replace monitor.

E09
DSC Shut Down

  1. Electro-cautery equipment used during self-test.
  2. Defective DSC.
  3. Defective monitor.
Note: This error requires monitor power down to reset error before each corrective action is performed.
  1. Isolate BIS system from possible noise sources or restart monitor when electro-cautery equipment not in use. (Note that self-test runs when the DSC is first connected.)
  2. Replace DSC.
  3. Replace monitor.

E10
Illegal DSC Type

  1. Incorrect DSC in use.
  2. Defective DSC.
  3. Defective monitor
  1. Connect the correct DSC.
  2. Replace DSC.
  3. Replace monitor.

E11
DSC Failed Self Test

  1. Electro-cautery equipment used during self-test.
  2. Defective DSC.
  3. Defective monitor.
  1. Isolate BIS system from possible noise sources or restart monitor when electro-cautery equipment not in use. (Note that self-test runs when the DSC is first connected.)
  2. Replace DSC.
  3. Replace monitor.

E12
Illegal Sensor Type

  1. Poor or contaminated connection between sensor and PIC.
  2. Defective sensor.
  3. Defective PIC.
  4. Defective DSC.
  1. Connect/clean connection between sensor and PIC.
  2. Replace the sensor.
  3. Replace PIC.
  4. Replace DSC.

E13, E15
Re-prep Sensor

  1. Incorrect sensor application.
  2. Poor sensor connections.
  3. Sensor Check fails
  4. Defective PIC.
  5. Defective DSC
  1. Read Instructions on sensor package and re-apply.
  2. Check sensor connections.
  3. Re-prep / replace sensor. Verify Sensor Check passes.
  4. Replace PIC.
  5. Replace DSC

E14
Sensor Not Connected

  1. Disconnected sensor
  2. Poor or contaminated connection between sensor and PIC.
  3. Disconnected PIC.
  4. Defective PIC.
  5. Defective DSC.
  6. Defective monitor
  1. Connect the Sensor
  2. Connect/clean connection between sensor and PIC.
  3. Connect the PIC.
  4. Replace PIC.
  5. Replace DSC.
  6. Replace monitor.

E15 (See E13)

   

E16
Last Sensor Check Failed

  1. At least one element of sensor has too high impedance, and EXIT pressed (before sensor check completes).
  2. Incorrect sensor application.
  3. Poor sensor connection.
  4. Defective PIC.
  5. Defective DSC.
  1. Verify Sensor Check passes.
  2. Read Instructions on sensor package and re-apply sensor.
  3. Check sensor connection.
  4. Replace PIC.
  5. Replace DSC.

E17, E19, E20, E21, E22, E23, E24
BIS Engine Comm. Error

Error in communication between BIS Engine and Host.
  1. Verify the software level.
  2. Replace monitor.

E18
BIS Engine Not Functional

Fatal BIS Engine error detected.
  1. Turn monitor off, then on again.
  2. Verify the software level.
  3. Replace monitor.

E19 - E24 (See E17)

   

 

EEG Signal (Errors 25 and 26):

Note: These messages relate to the condition of the EEG signal. This signal is often affected by environmental conditions that cause artifact (non-EEG signal). The messages below do not usually indicate a monitor system error; instead they serve as a notification of a patient or environmental condition.

Status Messages:

Causes:

Corrective Actions:

E25
Left Side SQI Bad

With 2-channel montage, left side signal quality is bad.
  1. Artifact is causing EEG information to be lost
  2. EMG Bar indicates electrical activity that may be interfering with EEG recognition.
  3. PIC is defective.
  4. DSC is defective.
Note: This message may occur as the result of artifact (non-EEG signal) such as those generated from motion (patient movement) or the presence of electro-cautery or other devices.
  1. If ARTIFACT label appears in EEG waveform box, attempt to identify and eliminate artifact source.
  2. If EMG is present, try to determine and eliminate cause.
  3. Verify Sensor Check passes. If not, replace PIC.
  4. Replace DSC

E26Right Side SQI Bad

With 2-channel montage, right side signal quality is bad.
  1. Artifact is causing EEG information to be lost
  2. EMG Bar indicates electrical activity that may be interfering with EEG recognition.
  3. PIC is defective.
  4. DSC is defective.
Note: This message may occur as the result of artifact (non-EEG signal) such as those generated from motion (patient movement) or the presence of electro-cautery or other devices.
  1. If ARTIFACT label appears in EEG waveform box, attempt to identify and eliminate artifact source.
  2. If EMG is present, try to determine and eliminate cause.
  3. Verify Sensor Check passes. If not, replace PIC.
  4. Replace DSC

E27
SQI Below 50

The Signal Quality Index is less than 50 and the numeric display changes from a solid to an outlined number.This may occur as the result of artifact such as those generated from motion or eyeblinks.
  1. Re-prep the sensor.
  2. Check impedance.

E28
SQI Too Low for BIS

The Signal Quality Index is unacceptable. The Primary Trend variable cannot be calculated and therefore the numeric display is blanked.
  1. Artifact, such as those generated by motion or eyeblinks, is causing loss of EEG recognition.
  2. EMG Bar indicates electrical activity that may be interfering with EEG recognition.
  3. PIC is defective.
  4. DSC is defective.
Note: This message may occur as the results of artifact (non-EEG signal) such as those generated from motion (patient movement or eyeblinks) or the presence of electro-cautery or other devices.
  1. If ARTIFACT label appears in EEG waveform box, attempt to identify and eliminate artifact source.
  2. If EMG bar is illuminated, attempt to determine and eliminate cause.
  3. Verify Sensor Check passes. If not, replace PIC.
  4. Replace DSC

E29
BIS < Low Alarm Limit

The BIS has fallen below the low alarm limit set by the user. The numeric display flashes. Take note of BIS at limit set by user.

E30
BIS > High Alarm Limit

The BIS has risen above the high alarm limit set by the user. The numeric display flashes. Take note of BIS at limit set by user.

E31
Isoelectric EEG Detected

No discernible EEG activity is detected for several minutes; SR=100.Note: This message notifies user of a flatline EEG. This is a normal condition when Sensor Simulator or Test Sensor is connected.
  1. Check the patient
  2. Check leads for proper connection and possible shorts.
  3. Verify Sensor Check passes.
  4. Verify DSC - Self-test
  5. Verify PIC Use Test Sensor or Sensor Simulator and Sensor Check.

E32
Review Data Lost

Data memory has been re-initialized.
  1. Turn the monitor off, then on again
  2. If error persists, monitor requires service.

E33
Operating On Battery Backup

The AC power has been lost and the monitor is running on the battery. The battery keeps the monitor operating for approximately 20 minutes (when the battery is fully charged).
  1. Restore the AC power.
  2. Verify power cord.
  3. Verify AC fuses are good.
  4. Replace monitor.

E34
Battery Power Low

There are only a few minutes of battery life left. Restore AC power to avoid automatic shutdown.

E35 - E39, E43, E44, E48 - E65, E69 E73, E75, E78 E80, E87- E89
Software Error

A software error has occurred. The monitor may stop operating.
  1. Turn the monitor off, then on again.
  2. Verify the software level.
  3. Replace monitor.

E40
Setup Memory Write Error

Error in erasing or writing Setup block of FLASH memory.
  1. Check Setup.
    Replace monitor.

E41
Real-Time Clock Write Error

Error in writing to Real-Time Clock RAM. Replace monitor.

E42
Cant Save Setup

Monitor is unable to write to Setup Memory when battery power is low.
  1. Restore AC power.
  2. Save settings again.
  3. Replace monitor.

E43, E44 (See E35)

   

E45
No Updates from BIS Engine

No data received from BIS Engine in last 10 seconds.
  1. Turn monitor off, then on again.
  2. Verify the software level.
  3. Replace monitor.

E46
BIS Engine Not Responding

BIS Engine has not responded to a command within 12 seconds.
  1. Turn monitor off, then on again.
  2. Verify the software level.
  3. Replace monitor.

E47, E74, E81
Data Memory Write Error

Error in erasing or writing Data Memory portion of FLASH memory. Replace monitor.

E48 E65 (See E35)

   

E66Printer Out of Paper

Printer is out of paper. Install the paper.

E67, E82, E85
Printer Error

  1. Printer is off-line.
  2. Print head is up.
  3. Printer is disconnected.
  4. Print head temperature out of range.
  5. Print head voltages out of range.
  6. Printer hardware failure.
  1. Put printer on-line.
  2. Close print head.
  3. Connect printer.
  4. Replace printer.

E68

Not used.  

E69 E73 (See E35)

   

E74 (See E47)

   

E75 (See E35)

   

E76
Illegal Sensor for Demo

A Demo Device has been connected to the serial port, but the PIC is connected to a sensor.
  1. Disconnect the sensor and connect the PIC to the Demo Device, or:
  2. Disconnect Demo Device.

E77
Simulator Connected

A Demo Device has been connected to the monitor. Note: This is a message rather than an error. No action is necessary.

E78 E80 (See E35)

   

E81 (See E47)

   

E82 (See E67)

   

E83
Printer Head is Up

The print head on the printer has been lifted.
  1. Return print head to the closed position.
  2. If the error persists, the printer requires service.

E84
Printer Disconnected

The printer is no longer connected to the A-2000.
  1. Check the connection.
  2. If the error persists, the printer requires service.

E85 (See E67)

   

E86
Printing Disabled

The A-2000 is running on battery backup. Note: Printer functions only on AC power. When power is restored, printing will again be available.

E87, E88, E89 (See E35)

   

E90
Printer Connected

The printer has been successfully connected to the A-2000. Note: This is a message rather than an error. No action is necessary.

E91
Event Marking Disabled

Data extraction is in progress. Note: Event marking will be re-enabled when event data extraction has been completed (less than 10 minutes).

E92

Not used  

E93
Sensor Ground Fault

Problem is detected relating to sensor ground element.
  1. Disconnect and examine sensor connection. Clean any contamination present.
  2. Replace sensor if necessary.
  3. Replace PIC.
  4. Replace DSC.

E94
Sensor Overcurrent

Sensor is using too much current.
  1. Disconnect and examine sensor connection. Clean any contamination.
  2. Replace sensor if necessary.
  3. Replace PIC.
  4. Replace DSC.

E95 E108
Invalid Sensor

The sensor is unusable for one of the following reasons:
  • E95 The sensor has been connected too many times.
  • E96-E102, E107 The sensor contains invalid information.
  • E103-E106, E108 The sensor information could not be read.
  1. Replace sensor.
  2. Replace PIC.
  3. Replace DSC.

Source: A-2000 XP Service Information Manual P/N 075-0014 Rev. 1.00.

 

For technical assistance contact:

In the U.S.A.:
Technical Service Department
Aspect Medical Systems, Inc.
Tel: 800-442-2051 (USA)
Tel: 617-559-7000 (USA)
Email: techsupport@aspectms.com

Outside the U.S.A.:
Technical Service
Aspect Medical Systems B.V.
Haagse Schouwweg 8b
NL - 2332 KG Leiden
Tel: +31-715 725 935

<< Back to Frequently Asked Questions | Back to Top

Copyright © 2009. Aspect Medical Systems. All Rights Reserved. Terms of Use | Privacy

080-0296 1.00